Nucleus =			1H;
Coils =				Q-Body, Base, Head, Posterior, (6)None;
Uniformity =			CLEAR;
FOV          AP (mm) =		230;
             RL (mm) =		182.955;
             FH (mm) =		129;
ACQ voxel size   AP (mm) =	0.898438;
             RL (mm) =		1.12443;
Slice thickness (mm) =		4;
Recon voxel size AP (mm) =	0.653409;
             RL (mm) =		0.653409;
Fold-over suppression =		no;
Reconstruction matrix =		352;
Acceleration =			AI CS-SENSE;
    reduction =			2;
    denoising =			system default;
    sharpen =			no;
Stacks =			1;
    type =			parallel;
    slices =			26;
    slice gap =			user defined;
          gap (mm) =		1;
    slice orientation =		transverse;
    fold-over direction =	RL;
    fat shift direction =	P;
Minimum number of packages =	3;
Slice scan order =		default;
O-MAR =				no;
PlanAlign =			no;
REST slabs =			0;
Interactive positioning =	no;
External control =		no;
Scan type =			Imaging;
Scan mode =			MS;
    technique =			IR;
Acquisition mode =		cartesian;
Fast Imaging mode =		TSE;
    shot mode =			multishot;
TSE factor =			41;
    startup echoes =		0;
    profile order =		linear;
    DRIVE =			no;
    ultrashort =		no;
    fid reduction =		default;
Echoes =			1;
    partial echo =		no;
TE =				user defined;
    (ms) =			150;
Refocusing control =		T2-optimized;
    min. angle  (deg) =		60;
    mid. angle (deg) =		90;
    max. angle (deg) =		120;
TR =				user defined;
    (ms) =			11000;
Halfscan =			no;
Water-fat shift =		maximum;
IR  delay (ms) =		2800;
    acquire during delay =	yes;
    dual =			no;
Shim =				default;
mDIXON =			no;
Fat suppression =		SPIR;
   strength =			strong;
   frequency offset =		default;
Water suppression =		no;
BB pulse =			no;
MTC =				no;
Silicone Only Sequence =	no;
Custom prepulse =		no;
SyntAc =			no;
Diffusion mode =		no;
T1 mapping =			no;
SAR mode =			moderate;
B1 mode =			default;
SAR allow first level =		yes;
PNS mode =			low;
Gradient mode =			default;
SofTone mode =			no;
Cardiac synchronization =	no;
Heart rate > 250 bpm =		no;
Respiratory compensation =	no;
Navigator respiratory comp =	no;
Flow compensation =		no;
Motion smoothing =		no;
NSA =				2;
    SMART =			no;
MRE enable =			no;
1D tracking =			no;
Manual start =			no;
Dynamic study =			no;
Arterial Spin labeling =	no;
Preparation phases =		auto;
Interactive F0 =		no;
Quick Survey =			default;
SmartPlan survey =		no;
B0 field map =			no;
B1 field map =			no;
MIP/MPR =			no;
Images =			M, (3)no;
Autoview image =		M;
Reference tissue =		White matter;
Recon compression =		No;
Preset window contrast =	soft;
Reconstruction mode =		real time;
Hardcopy protocol =		no;
Image filter =			system default;
Uniformity correction =		no;
Geometry correction =		default;
MR Therapy protocol =		no;
Stack Offc. AP (P=+mm) =	-41.4638;
            RL (L=+mm) =	-3.68458;
            FH (H=+mm) =	37.7788;
      Ang.  AP (deg) =		-1.40908;
            RL (deg) =		5.19894;
            FH (deg) =		-1.00226;
      Free rotatable =		no;