Nucleus =			1H;
Uniformity =			CLEAR;
FOV          FH (mm) =		250;
             AP (mm) =		250;
             RL (mm) =		182.56;
ACQ voxel size   FH (mm) =	1.12;
             AP (mm) =		1.12108;
             RL (mm) =		1.12;
Recon voxel size FH (mm) =	0.744048;
             AP (mm) =		0.744048;
             RL (mm) =		0.56;
Fold-over suppression =		no;
Slice oversampling =		user defined;
    oversample factor =		1;
Reconstruction matrix =		336;
Acceleration =			CS-SENSE;
    reduction =			8.5;
    denoising =			weak;
Stacks =			1;
    slices =			326;
    slice orientation =		sagittal;
    fold-over direction =	AP;
    fat shift direction =	F;
Multi-chunk =			no;
O-MAR =				no;
PlanAlign =			no;
REST slabs =			0;
Interactive positioning =	no;
Scan type =			Imaging;
Scan mode =			3D;
    technique =			IR;
Acquisition mode =		cartesian;
Fast Imaging mode =		TSE;
    3D VIEW =			Brain FLAIR;
    shot mode =			multishot;
TSE factor =			167;
    startup echoes =		6;
    profile order =		linear;
    turbo direction =		Y;
    DRIVE =			no;
    fid reduction =		default;
Echoes =			1;
    partial echo =		no;
TE =				user defined;
    (ms) =			340;
Refocusing control =		constant;
    angle (deg) =		40;
    bright fat reduction =	no;
TR =				user defined;
    (ms) =			4800;
Halfscan =			no;
Water-fat shift =		maximum;
IR  delay (ms) =		1650;
    dual =			no;
    power =			1;
Shim =				default;
mDIXON =			no;
Fat suppression =		SPIR;
   strength =			strong;
   frequency offset =		default;
Grad Rev Fat suppr =		no;
Water suppression =		no;
BB pulse =			no;
MTC =				no;
T2prep =			yes;
    echo time (ms) =		125;
    refocusing pulses =		4;
Custom prepulse =		no;
MDME =				no;
Diffusion mode =		no;
T1 mapping =			no;
SAR mode =			high;
B1 mode =			default;
SAR allow first level =		yes;
PNS mode =			high;
Gradient mode =			maximum;
SofTone mode =			no;
Cardiac synchronization =	no;
Heart rate > 250 bpm =		no;
Respiratory compensation =	no;
Navigator respiratory comp =	no;
Flow compensation =		no;
Motion smoothing =		no;
NSA =				2;
    partial NSA =		no;
    SMART =			no;
MRE enable =			no;
CENTRA =			no;
Manual start =			no;
Dynamic study =			no;
Arterial Spin labeling =	no;
Preparation phases =		auto;
Interactive F0 =		no;
Quick Survey =			default;
SmartPlan survey =		no;
B0 field map =			no;
B1 field map =			no;
MIP/MPR =			no;
Images =			M, (3)no;
Autoview image =		M;
Reference tissue =		Grey matter;
Recon compression =		No;
Preset window contrast =	soft;
Reconstruction mode =		real time;
Save raw data =			no;
Hardcopy protocol =		no;
Image filter =			system default;
Uniformity correction =		yes;
Geometry correction =		default;
MR Therapy protocol =		no;
Elliptical k-space shutter =	default;
Stack Offc. AP (P=+mm) =	-25.7871;
            RL (L=+mm) =	-2.17779;
            FH (H=+mm) =	-48.8904;
      Ang.  AP (deg) =		-1.48026;
            RL (deg) =		0.737581;
            FH (deg) =		-0.176252;
      Free rotatable =		no;
Act. slice gap (mm) =		-0.56;
Act. TR/TI (ms) =		4800 / 1650;
Act. TE (ms) =			340;
Min. TR/TI (ms) =		2446 / 50;
Total scan duration =		03:16.8;
Rel. SNR =			1;
ACQ matrix M x P =		224 x 223;
ACQ voxel MPS (mm) =		1.12 / 1.12 / 1.12;
REC voxel MPS (mm) =		0.74 / 0.74 / 0.56;
Scan percentage (%) =		99.5536;
WFS (pix) / BW (Hz) =		0.654 / 664.3;
TSE es / shot (ms) =		3.7 / 647;
TEeff / TEequiv (ms) =		340 / 142 ;
Head SAR =			<  15 %;
Whole body SAR / level =	< 0.1 W/kg / normal;
SED =				  0.0 kJ/kg;
Max B1+rms =			0.93 uT;
PNS / level =			76 % / normal;
dB/dt =				103.4 T/s;
Sound Pressure Level (dB) =	11.3877;
Boil-off (hPa/h) =		1.07176;
Max Boil-off (hPa/h) =		8.15102;
Boil-off (hPa) =		0.0585894;